Αρχειοθήκη ιστολογίου

Πέμπτη 4 Μαρτίου 2021

Novel C‐arm based planning spine surgery robot proved in a porcine model and quantitative accuracy assessment methodology

xlomafota13 shared this article with you from Inoreader

Abstract

Background

We assessed pedicle screw accuracy utilizing a novel navigation‐based spine surgery robotic system by comparing planned pathways with placed pathways in a porcine model.

Methods

We placed three mini screws per vertebra for accuracy evaluation and used a reference frame for registration in four pigs (46 screws in 23 vertebrae). We planned screw paths and performed screw insertion under robot guidance. Using C‐arm and CT images, we evaluated accuracy by comparing the 3D distance of the placed screw head/tip from the planned screw head/tip and 3D angular offset.

Results

Mean registration deviation between the preoperative 3D space (C‐arm) and postoperative CT scans was 0.475 ± 0.119 mm. The average offset from preoperative plan to final placement was 4.8 ± 2.0 mm from the head (tail), 5.3 ± 2.3 mm from the tip and 3.9 ± 2.4 degrees of angulation.

Conclusions

Our spine surgery robot showed good accuracy in executing an intended planned trajectory and screw path. This faster and more accurate robotic system will be applied in future studies, first in cadavers and subsequently in the clinical field.

View on the web

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου