Αρχειοθήκη ιστολογίου

Τρίτη 5 Σεπτεμβρίου 2017

Impact of Intentional Overload on Joint Stability of Internal Implant-Abutment Connection System with Different Diameter

Abstract

Purpose

To evaluate the axial displacement of the implant-abutment assembly of different implant diameter after static and cyclic loading of overload condition.

Materials and Methods

An internal conical connection system with three diameters (Ø 4.0, 4.5, and 5.0) applying identical abutment dimension and the same abutment screw was evaluated. Axial displacement of abutment and reverse torque loss of abutment screw were evaluated under static and cyclic loading conditions. Static loading test groups were subjected to vertical static loading of 250, 400, 500, 600, 700, and 800 N consecutively. Cyclic loading test groups were subjected to 500 N cyclic loading to evaluate the effect of excessive masticatory loading. After abutment screw tightening for 30 Ncm, axial displacement was measured upon 1, 3, 10, and 1,000,000 cyclic loadings of 500 N. Repeated-measure ANOVA and 2-way ANOVA were used for statistical analysis (α = 0.05).

Results

The increasing magnitude of vertical load and thinner wall thickness of implant increased axial displacement of abutment and reverse torque loss of abutment screw (p < 0.05). Implants in the Ø 5.0 diameter group demonstrated significantly low axial displacement, and reverse torque loss after static loading than Ø 4.0 and Ø 4.5 diameter groups (p < 0.05). In the cyclic loading test, all diameter groups of implant showed significant axial displacement after 1 cycle of loading of 500 N (p < 0.05). There was no significant axial displacement after 3, 10, or 1,000,000 cycles of loading (p = 0.603).

Conclusions

Implants with Ø 5.0 diameter demonstrated significantly low axial displacement and reverse torque loss after the cyclic and static loading of overload condition.



from #ORL-AlexandrosSfakianakis via ola Kala on Inoreader http://ift.tt/2gHggYF

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου