Blind intubation of anaesthetised children with supraglottic airway devices AmbuAura-i and Air-Q cannot be recommended: A randomised controlled trial.

http:--pt.wkhealth.com-pt-pt-core-templa Related Articles

Blind intubation of anaesthetised children with supraglottic airway devices AmbuAura-i and Air-Q cannot be recommended: A randomised controlled trial.

Eur J Anaesthesiol. 2015 Sep;32(9):631-9

Authors: Kleine-Brueggeney M, Nicolet A, Nabecker S, Seiler S, Stucki F, Greif R, Theiler L

Abstract
BACKGROUND: Paediatric supraglottic airway devices AmbuAura-i and Air-Q were designed as conduits for tracheal intubation. Although fibreoptic-guided intubation has proved successful, blind intubation as a rescue technique has never been evaluated.
OBJECTIVE: Evaluation of blind intubation through AmbuAura-i and Air-Q. On the basis of fibreoptic view data, we hypothesised that the success rate with the AmbuAura-i would be higher than with the Air-Q.
DESIGN: A randomised controlled trial.
SETTING: University Childrens' Hospital; September 2012 to July 2014.
PATIENTS: Eighty children, American Society of Anesthesiologists (ASA) class I to III, weight 5 to 50 kg.
INTERVENTIONS: Tracheal intubation was performed through the randomised device with the tip of a fibrescope placed inside and proximal to the tip of the tracheal tube. This permitted sight of tube advancement, but without fibreoptic guidance (visualised blind intubation).
MAIN OUTCOME MEASURES: Primary outcome was successfully visualised blind intubation; secondary outcomes included supraglottic airway device success, insertion times, airway leak pressure, fibreoptic view and adverse events.
RESULTS: Personal data did not differ between groups. In contrast to our hypothesis, blind intubation was possible in 15% with the Air-Q and in 3% with the AmbuAura-i [95% confidence interval (95% CI) 6 to 31 vs. 0 to 13%; P = 0.057]. First attempt supraglottic airway device insertion success rates were 95% (Air-Q) and 100% (AmbuAura-i; 95% CI 83 to 99 vs. 91 to 100; P = 0.49). Median leak pressures were 18 cmH2O (Air-Q) and 17 cmH2O [AmbuAura-i; interquartile range (IQR) 14 to 18 vs. 14 to 19 cmH2O; P = 0.66]. Air-Q insertion was slower (27 vs. 19 s, P < 0.001). There was no difference in fibreoptic view, or adverse events (P > 0.05). In one child (Air-Q size 1.5, tube size 3.5), the tube dislocated during device removal.
CONCLUSION: Ventilation with both devices is reliable, but success of blind intubation is unacceptably low and cannot be recommended for elective or rescue purposes. If intubation through a paediatric supraglottic airway device is desired, we suggest that fibreoptic guidance is used.
TRIAL REGISTRATION: Clinicaltrials.gov identifier: NCT01692522.

PMID: 26061874 [PubMed - indexed for MEDLINE]